Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Tektronix 2445 (1 year) CAL VER #1 /5011 DATE: 13-Mar-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 43 NUMBER OF LINES: 372 CONFIGURATION: Tektronix CG5011 ============================================================================= # # Source: # Tektronix 2445 Oscilloscope Service Manual 070-3829-00, Revised Jan 1985 # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default CG 5011 and 6060B accuracy files contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- N P J F W 1.002 ASK+ R D 1.003 HEAD CH [MEM2] VOLTS/DIV, DELTA V, TRIGGER LEVEL ACCURACIES 1.004 RSLT =CH [MEM2] VOLTS/DIV, DELTA V, TRIGGER LEVEL ACCURACIES 1.005 DISP Connect the CG5011 Pulse Head to UUT Channel [MEM2]. 1.006 DISP VOLTS/DIV CH [MEM2] : 2mV 1.007 5011 10mVpp 1kH -D S 1.008 DISP Use the VERTICAL POSITION control to set the bottom of 1.008 DISP the signal 2 divisions below graticule center. 1.008 DISP Rotate the REF or DLY POS control to align the 1.008 DISP reference cursor with the bottom of the waveform. 1.008 DISP Rotate the control to align the cursor 1.008 DISP with the top of the signal display. 1.008 DISP Vertical deflection should be 4.9 to 5.1 divisions. 1.008 DISP Delta Volts Readout should be 9.81 to 10.20 mV. 1.009 EVAL Is the {Vertical Deflection 4.9 to 5.1 divisions}? 2.001 5011 10mVpp 1kH -D S 2.002 EVAL Is the {Delta Volts Readout 9.81 to 10.20 mV}? 3.001 5011 10mVpp 1kH -D S 3.002 DISP Set the TRIGGER LEVEL control at the most positive 3.002 DISP voltage that produces a barely triggered, jittering 3.002 DISP display for each position of the SLOPE switch. 3.002 DISP 3.002 DISP A Trigger Level Reading should be 8.0 to 12.0 mV. 3.003 EVAL Is the {A Trigger Level Reading 8.0 to 12.0 mV}? 4.001 5011 10mVpp 1kH -D S 4.002 DISP Set the TRIGGER LEVEL control at the most negative 4.002 DISP voltage that produces a barely triggered, jittering 4.002 DISP display for each position of the SLOPE switch. 4.003 DISP 4.003 DISP A Trigger Level Reading should be +1.7 to -1.7 mV. 4.004 EVAL Is the {A Trigger Level Reading +1.7 to -1.7 mV}? 5.001 5011 10mVpp 1kH -D S 5.002 DISP Set the TRIGGER LEVEL for a stable display. 5.002 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE to 5.002 DISP TRIG AFT DLY. 5.002 DISP 5.002 DISP Adjust REF or DLY POS control for a delay reading 5.002 DISP of 0.000ms. 5.003 DISP Set the TRIGGER LEVEL control at the most positive 5.003 DISP voltage that produces an intensified point on the 5.003 DISP waveform display for each position of the SLOPE switch. 5.003 DISP 5.003 DISP B Trigger Level Reading should be 8.0 to 12.0 mV. 5.004 EVAL Is the {B Trigger Level Reading 8.0 to 12.0 mV}? 6.001 5011 10mVpp 1kH -D S 6.002 DISP Set the TRIGGER LEVEL control at the most negative 6.002 DISP voltage that produces an intensified point on the 6.002 DISP waveform display for each position of the SLOPE switch. 6.002 DISP 6.002 DISP B Trigger Level Reading should be +1.7 to -1.7 mV. 6.003 EVAL Is the {B Trigger Level Reading +1.7 to -1.7 mV}? 7.001 DISP Push the B SEC/DIV knob in. 7.002 DISP VOLTS/DIV CH [MEM2] : 5mV 7.003 5011 20mVpp 1kH -D S 7.004 DISP Press and release the V pushbutton to reobtain 7.004 DISP the V readout display. 7.004 DISP Rotate the REF or DLY POS control to align the 7.004 DISP reference cursor with the bottom of the waveform. 7.004 DISP Rotate the control to align the cursor 7.004 DISP with the top of the signal display. 7.004 DISP Vertical deflection should be 3.92 to 4.08 div. 7.004 DISP Delta Volts Readout should be 19.6 to 20.4mV. 7.005 EVAL Is the {Vertical Deflection 3.92 to 4.08 divisions}? 8.001 5011 20mVpp 1kH -D S 8.002 EVAL Is the {Delta Volts Readout 19.6 to 20.4 mV}? 9.001 5011 20mVpp 1kH -D S 9.002 DISP Set the TRIGGER LEVEL control at the most positive 9.002 DISP voltage that produces a barely triggered, jittering 9.002 DISP display for each position of the SLOPE switch. 9.002 DISP 9.002 DISP A Trigger Level Reading should be 16.8 to 23.2 mV. 9.003 EVAL Is the {A Trigger Level Reading 16.8 to 23.2 mV}? 10.001 5011 20mVpp 1kH -D S 10.002 DISP Set the TRIGGER LEVEL control at the most negative 10.002 DISP voltage that produces a barely triggered, jittering 10.002 DISP display for each position of the SLOPE switch. 10.002 DISP 10.002 DISP A Trigger Level Reading should be +2.6 to -2.6 mV. 10.003 EVAL Is the {A Trigger Level Reading +2.6 to -2.6 mV}? 11.001 5011 20mVpp 1kH -D S 11.002 DISP Set the TRIGGER LEVEL for a stable display. 11.002 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE to 11.002 DISP TRIG AFT DLY. 11.003 DISP Adjust REF or DLY POS control for a delay reading 11.003 DISP of 0.000ms. 11.004 DISP Set the TRIGGER LEVEL control at the most positive 11.004 DISP voltage that produces an intensified point on the 11.004 DISP waveform display for each position of the SLOPE switch. 11.004 DISP 11.004 DISP B Trigger Level Reading should be 16.8 to 23.2 mV. 11.005 EVAL Is the {B Trigger Level Reading 16.8 to 23.2 mV}? 12.001 5011 20mVpp 1kH -D S 12.002 DISP Set the TRIGGER LEVEL control at the most negative 12.002 DISP voltage that produces an intensified point on the 12.002 DISP waveform display for each position of the SLOPE switch. 12.002 DISP 12.002 DISP B Trigger Level Reading should be +2.6 to -2.6 mV. 12.003 EVAL Is the {B Trigger Level Reading +2.6 to -2.6 mV}? 13.001 DISP Push the B SEC/DIV knob in. 13.002 DISP VOLTS/DIV CH [MEM2] : 10mV 13.003 5011 50mVpp 1kH -D S 13.004 DISP Press and release the V pushbutton to reobtain 13.004 DISP the V readout display. 13.005 DISP Use the VERTICAL POSITION control to set the bottom of 13.005 DISP the signal 2 divisions below graticule center. 13.005 DISP Rotate the REF or DLY POS control to align the 13.005 DISP reference cursor with the bottom of the waveform. 13.005 DISP Rotate the control to align the cursor 13.005 DISP with the top of the signal display. 13.005 DISP Vertical deflection should be 4.9 to 5.1 divisions. 13.005 DISP Delta Volts Readout should be 49.0 to 50.9 mV. 13.006 EVAL Is the {Vertical Deflection 4.90 to 5.10 divisions}? 14.001 5011 50mVpp 1kH -D S 14.002 EVAL Is the {Delta Volts Readout 49.0 to 50.9 mV}? 15.001 5011 50mVpp 1kH -D S 15.002 DISP Set the TRIGGER LEVEL control at the most positive 15.002 DISP voltage that produces a barely triggered, jittering 15.002 DISP display for each position of the SLOPE switch. 15.002 DISP 15.002 DISP A Trigger Level Reading should be 44 to 56 mV. 15.003 EVAL Is the {A Trigger Level Reading 44 to 56 mV}? 16.001 5011 50mVpp 1kH -D S 16.002 DISP Set the TRIGGER LEVEL control at the most negative 16.002 DISP voltage that produces a barely triggered, jittering 16.002 DISP display for each position of the SLOPE switch. 16.002 DISP 16.002 DISP A Trigger Level Reading should be +4.5 to -4.5mV 16.003 EVAL Is the {A Trigger Level Reading +4.5 to -4.5 mV}? 17.001 5011 50mVpp 1kH -D S 17.002 DISP Set the TRIGGER LEVEL for a stable display. 17.002 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE to 17.002 DISP TRIG AFT DLY. 17.003 DISP Adjust REF or DLY POS control for a delay reading 17.003 DISP of 0.000ms. 17.004 DISP Set the TRIGGER LEVEL control at the most positive 17.004 DISP voltage that produces an intensified point on the 17.004 DISP waveform display for each position of the SLOPE switch. 17.004 DISP 17.004 DISP B Trigger Level Reading should be 44 to 56 mV. 17.005 EVAL Is the {B Trigger Level Reading 44 to 56 mV}? 18.001 5011 50mVpp 1kH -D S 18.002 DISP Set the TRIGGER LEVEL control at the most negative 18.002 DISP voltage that produces an intensified point on the 18.002 DISP waveform display for each position of the SLOPE switch. 18.002 DISP 18.002 DISP B Trigger Level Reading should be +4.5 to -4.5 mV. 18.003 EVAL Is the {B Trigger Level Reading +4.5 to -4.5 mV}? 19.001 DISP Push the B SEC/DIV knob in. 19.002 DISP VOLTS/DIV CH [MEM2] : 20mV 19.003 5011 0.1Vpp 1kH -D S 19.004 DISP Press and release the V pushbutton to reobtain 19.004 DISP the V readout display. 19.005 DISP Rotate the REF or DLY POS control to align the 19.005 DISP reference cursor with the bottom of the waveform. 19.005 DISP Rotate the control to align the cursor 19.005 DISP with the top of the signal display. 19.005 DISP Vertical deflection should be 4.9 to 5.1 divisions. 19.005 DISP Delta Volts Readout should be 98.1 to 102.0 mV. 19.006 EVAL Is the {Vertical Deflection 4.90 to 5.10 divisions}? 20.001 5011 0.1Vpp 1kH -D S 20.002 EVAL Is the {Delta Volts Readout 98.1 to 102.0 mV}? 21.001 5011 0.1Vpp 1kH -D S 21.002 DISP Set the TRIGGER LEVEL control at the most positive 21.002 DISP voltage that produces a barely triggered, jittering 21.002 DISP display for each position of the SLOPE switch. 21.002 DISP 21.002 DISP A Trigger Level Reading should be 89 to 111 mV. 21.003 EVAL Is the {A Trigger Level Reading 89 to 111 mV}? 22.001 5011 0.1Vpp 1kH -D S 22.002 DISP Set the TRIGGER LEVEL control at the most negative 22.002 DISP voltage that produces a barely triggered, jittering 22.002 DISP display for each position of the SLOPE switch. 22.002 DISP 22.002 DISP A Trigger Level Reading should be +8.0 to -8.0mV 22.003 EVAL Is the {A Trigger Level Reading +8.0 to -8.0 mV}? 23.001 5011 0.1Vpp 1kH -D S 23.002 DISP Set the TRIGGER LEVEL for a stable display. 23.002 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE to 23.002 DISP TRIG AFT DLY. 23.003 DISP Adjust REF or DLY POS control for a delay reading 23.003 DISP of 0.000ms. 23.004 DISP Set the TRIGGER LEVEL control at the most positive 23.004 DISP voltage that produces an intensified point on the 23.004 DISP waveform display for each position of the SLOPE switch. 23.004 DISP 23.004 DISP B Trigger Level Reading should be 89 to 111 mV. 23.005 EVAL Is the {B Trigger Level Reading 89 to 111 mV}? 24.001 5011 0.1Vpp 1kH -D S 24.002 DISP Set the TRIGGER LEVEL control at the most negative 24.002 DISP voltage that produces an intensified point on the 24.002 DISP waveform display for each position of the SLOPE switch. 24.002 DISP 24.002 DISP B Trigger Level Reading should be +8.0 to -8.0 mV. 24.003 EVAL Is the {B Trigger Level Reading +8.0 to -8.0 mV}? 25.001 DISP Push the B SEC/DIV knob in. 25.002 DISP VOLTS/DIV CH [MEM2] : 50mV 25.003 5011 0.2Vpp 1kH -D S 25.004 DISP Press and release the V pushbutton to reobtain 25.004 DISP the V readout display. 25.005 DISP Use the VERTICAL POSITION control to set the bottom of 25.005 DISP the signal 2 divisions below graticule center. 25.005 DISP Rotate the REF or DLY POS control to align the 25.005 DISP reference cursor with the bottom of the waveform. 25.005 DISP Rotate the control to align the cursor 25.005 DISP with the top of the signal display. 25.005 DISP Vertical deflection should be 3.92 to 4.08 divisions. 25.005 DISP Delta Volts Readout should be 196 to 204 mV. 25.006 EVAL Is the {Vertical Deflection 3.92 to 4.08 divisions}? 26.001 5011 0.2Vpp 1kH -D S 26.002 EVAL Is the {Delta Volts Readout 196 to 204 mV}? 27.001 5011 0.2Vpp 1kH -D S 27.002 DISP Set the TRIGGER LEVEL control at the most positive 27.002 DISP voltage that produces a barely triggered, jittering 27.002 DISP display for each position of the SLOPE switch. 27.002 DISP 27.002 DISP A Trigger Level Reading should be 177 to 223 mV. 27.003 EVAL Is the {A Trigger Level Reading 177 to 223 mV}? 28.001 5011 0.2Vpp 1kH -D S 28.002 DISP Set the TRIGGER LEVEL control at the most negative 28.002 DISP voltage that produces a barely triggered, jittering 28.002 DISP display for each position of the SLOPE switch. 28.002 DISP 28.002 DISP A Trigger Level Reading should be +17.5 to -17.5 mV. 28.003 EVAL Is the {A Trigger Level Reading +17.5 to -17.5 mV}? 29.001 5011 0.2Vpp 1kH -D S 29.002 DISP Set the TRIGGER LEVEL for a stable display. 29.002 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE to 29.002 DISP TRIG AFT DLY. 29.003 DISP Adjust REF or DLY POS control for a delay reading 29.003 DISP of 0.000ms. 29.004 DISP Set the TRIGGER LEVEL control at the most positive 29.004 DISP voltage that produces an intensified point on the 29.004 DISP waveform display for each position of the SLOPE switch. 29.005 DISP 29.005 DISP B Trigger Level Reading should be 177 to 223 mV. 29.006 EVAL Is the {B Trigger Level Reading 177 to 223 mV}? 30.001 5011 0.2Vpp 1kH -D S 30.002 DISP Set the TRIGGER LEVEL control at the most negative 30.002 DISP voltage that produces an intensified point on the 30.002 DISP waveform display for each position of the SLOPE switch. 30.002 DISP 30.002 DISP B Trigger Level Reading should be +17.5 to -17.5 mV. 30.003 EVAL Is the {B Trigger Level Reading +17.5 to -17.5 mV}? 31.001 DISP Push the B SEC/DIV knob in. 31.002 DISP VOLTS/DIV CH [MEM2] : .1V 31.003 5011 0.5Vpp 1kH -D S 31.004 DISP Press and release the V pushbutton to reobtain 31.004 DISP the V readout display. 31.005 DISP Rotate the REF or DLY POS control to align the 31.005 DISP reference cursor with the bottom of the waveform. 31.005 DISP Rotate the control to align the cursor 31.005 DISP with the top of the signal display. 31.005 DISP Vertical deflection should be 4.9 to 5.1 divisions. 31.006 DISP Delta Volts Readout should be 490 to 509 mV. 31.007 EVAL Is the {Vertical Deflection 4.90 to 5.10 divisions}? 32.001 5011 0.5Vpp 1kH -D S 32.002 EVAL Is the {Delta Volts Readout 490 to 509 mV}? 33.001 5011 0.5Vpp 1kH -D S 33.002 DISP Set the TRIGGER LEVEL control at the most positive 33.002 DISP voltage that produces a barely triggered, jittering 33.002 DISP display for each position of the SLOPE switch. 33.002 DISP 33.002 DISP A Trigger Level Reading should be 0.449 to 0.551 V. 33.003 EVAL Is the {A Trigger Level Reading 0.449 to 0.551 V}? 34.001 5011 0.5Vpp 1kH -D S 34.002 DISP Set the TRIGGER LEVEL control at the most negative 34.002 DISP voltage that produces a barely triggered, jittering 34.002 DISP display for each position of the SLOPE switch. 34.002 DISP 34.002 DISP A Trigger Level Reading should be +0.036 to -0.036 V. 34.003 EVAL Is the {A Trigger Level Reading +0.036 to -0.036 V}? 35.001 5011 0.5Vpp 1kH -D S 35.002 DISP Set the TRIGGER LEVEL for a stable display. 35.002 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE to 35.002 DISP TRIG AFT DLY. 35.003 DISP Adjust REF or DLY POS control for a delay reading 35.003 DISP of 0.000ms. 35.004 DISP Set the TRIGGER LEVEL control at the most positive 35.004 DISP voltage that produces an intensified point on the 35.004 DISP waveform display for each position of the SLOPE switch. 35.004 DISP 35.004 DISP B Trigger Level Reading should be 89 to 111 mV. 35.005 EVAL Is the {B Trigger Level Reading 0.449 to 0.551 V}? 36.001 5011 0.5Vpp 1kH -D S 36.002 DISP Set the TRIGGER LEVEL control at the most negative 36.002 DISP voltage that produces an intensified point on the 36.002 DISP waveform display for each position of the SLOPE switch. 36.002 DISP 36.002 DISP B Trigger Level Reading should be +0.036 to -0.036V. 36.003 EVAL Is the {B Trigger Level Reading +0.036 to -0.036 V}? 37.001 DISP Push the B SEC/DIV knob in. 37.002 DISP VOLTS/DIV CH [MEM2] : .2V 37.003 5011 1.0Vpp 1kH -D S 37.004 DISP Press and release the V pushbutton to reobtain 37.004 DISP the V readout display. 37.005 DISP Rotate the REF or DLY POS control to align the 37.005 DISP reference cursor with the bottom of the waveform. 37.005 DISP Rotate the control to align the cursor 37.005 DISP with the top of the signal display. 37.005 DISP Vertical deflection should be 4.9 to 5.1 divisions. 37.005 DISP Delta Volts Readout should be 0.981 to 1.020 V. 37.006 EVAL Is the {Vertical Deflection 4.90 to 5.10 divisions}? 38.001 5011 1.0Vpp 1kH -D S 38.002 EVAL Is the {Delta Volts Readout 0.981 to 1.020 V}? 39.001 5011 1.0Vpp 1kH -D S 39.002 DISP Set the TRIGGER LEVEL control at the most positive 39.002 DISP voltage that produces a barely triggered, jittering 39.002 DISP display for each position of the SLOPE switch. 39.002 DISP 39.002 DISP A Trigger Level Reading should be 0.9 to 1.1 V. 39.003 EVAL Is the {A Trigger Level Reading 0.90 to 1.10 V}? 40.001 5011 1.0Vpp 1kH -D S 40.002 DISP Set the TRIGGER LEVEL control at the most negative 40.002 DISP voltage that produces a barely triggered, jittering 40.002 DISP display for each position of the SLOPE switch. 40.002 DISP 40.002 DISP A Trigger Level Reading should be +0.07 to -0.07 V. 40.003 EVAL Is the {A Trigger Level Reading +0.07 to -0.07 V}? 41.001 5011 1.0Vpp 1kH -D S 41.002 DISP Set the TRIGGER LEVEL for a stable display. 41.002 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE to 41.002 DISP TRIG AFT DLY. 41.003 DISP Adjust REF or DLY POS control for a delay reading 41.003 DISP of 0.000ms. 41.004 DISP Set the TRIGGER LEVEL control at the most positive 41.004 DISP voltage that produces an intensified point on the 41.004 DISP waveform display for each position of the SLOPE switch. 41.005 DISP 41.005 DISP B Trigger Level Reading should be 0.9 to 1.1 V. 41.006 EVAL Is the {B Trigger Level Reading 0.90 to 1.10 V}? 42.001 5011 1.0Vpp 1kH -D S 42.002 DISP Set the TRIGGER LEVEL control at the most negative 42.002 DISP voltage that produces an intensified point on the 42.002 DISP waveform display for each position of the SLOPE switch. 42.002 DISP 42.002 DISP B Trigger Level Reading should be +0.07 to -0.07 V. 42.003 EVAL Is the {B Trigger Level Reading +0.07 to -0.07 V}? 43.001 DISP Push the B SEC/DIV knob in. 43.002 END